Welcome to ECOL - Join Free - Sign In
QS 30T Load Cell
QS 30T Load Cell
Supplier Info
[China Supplier]
Contact Person : Ms. Huang Anny
Tel : 86-595-85712822
Fax : 86-595-85727357
Product Detail
*.Construction with steel ball and cup *.High precision and reliability supper stability *.easy installation

Load Cell technician:

Capacity

t

10/20/25/30/40SE/40/50

Accuracy

C1

C2

C3

A5M

B10M

Approvals

OIML R60 C3

Maximum number ofverification intervals

n max

1000

2000

3000

5000

10000

Minimum load cell verification interval

v min

Emax/3000

Emax/5000

Emax/7500

Emax/15000

Emax/10000

Combined error

(%FS)

≤±0.050

≤±0.030

≤±0.020

≤±0.026

≤±0.050

Creep

(%FS/30min)

≤±0.038

≤±0.024

≤±0.016

≤±0.017

≤±0.040

Temperature effect on sensitivity

(%FS/10°C)

≤±0.028

≤±0.017

≤±0.011

≤±0.013

≤±0.040

Temperature effect on zero

(%FS/10°C)

≤±0.047

≤±0.023

≤±0.015

≤±0.014

≤±0.020

Output sensitivity

(mv/v)

2.0±0.002

2.3±0.002

2.4±0.002

1.9±0.002

1.8±0.002

Input resistance

(Ω)

700 ±7

775 ±5

750 ±7

Output resistance

(Ω)

700 ±7

Insulation resistance

(MΩ)

≥ 5000(50VDC)

Zero balance

(%FS)

1.5

Temperature, compensated

(°C)

-10~+40

Temperature,allowed

(°C)

-35~+65

Excitation, Recommended

(V)

5~12(DC)

Excitation,Max

(V)

18(DC)

Safe overload

(%FS)

150

Ultimate overload

(%FS)

300