Air Sampler (Microbial Air Monitoring System)
The Huankai Air Sampler is a high specification instrument for monitoring the level of microbial contamination in pharmaceutical, clinical and food environments.
95% efficiency for 0.8 to 19.0 micron particle size.
CE Qualified
EASY TO USE
The simple keypad makes menu navigation easy, with all sampling information
clearly displayed on a liquid crystal display. The unit is light enough to hold
while sampling, or can be mounted on a tripod
FLEXIBLE
The sampler has simple set user-defined functions that offer flexible sampling volumes, delayed start. Various accessories are available for different applications, including high quality aviation aluminum heads, tripods and carry cases.
COLLECTION EFFICIENCY
Validation work according to BS EN ISO 14698-1:2003 Annex B, as carried out by an independent laboratory, has shown the air sampler to have a physical efficiency of 95% for particles, 0.8 to 19.0 micron in size.
DATA HANDLING
Each sampling run generates a unique record which is stored within the unit
(holds up to 512 records). The data is easily accessed and read on the liquid crystal display. Alternatively, the Huankai Air Sampler software allows data to be downloaded to a computer for reporting, printing and trend analysis.
CALIBRATION
The instrument is supplied with a traceable calibration certificate, valid for
twelve months. There is no requirement for additional calibration between
sampling head changes. A regular recalibration is available from Huankai.
SERVICING AND TECHNICAL SUPPORT
The Huankai Air Sampler comes with a twelve-month manufacturer's
warranty. Huankai provides service, calibration and full technical support.
Contact your local representative for details.
Technical Specifications
Principle: Air is aspirated (at a fixed speed for variable times) through
small holes in the instrument's head. The resulting laminar air flow is
directed onto the surface of an agar plate. At the end of the sampling
cycle, the plate is removed and incubated and the visible colonies are
then counted for an assessment of the level of contamination.
Collection Efficiency: Physical efficiency was determined using
Bacillus subtilis spores and a membrane filter reference method
(BS EN ISO 14698-1:2003 Annex B).
Nominal Air Flow Rate: 100 litres of air per minute.
Impact Speed: ≈20m/sec.
Air Sample Volumes: 10-6000 litres; up to 100 volumes can be selected and stored in the instrument (USER MODE). The maximum volume of air
per cycle is 6000 litres.
Size: 120 x 170 x 325mm.
Weight: 2.68kg.
Construction:
Sampling Head: high quality aviation aluminum
Fan: antistatic plastic resin.
Housing: high quality aviation aluminum.
Battery Pack: Rechargeable Ni/MH (without memory
effect). Battery charging takes only 6 hours. After
charging, the battery will last up to 20 hours.
Ordering Details:
Air Sampler:
The Huankai Air Sampler comes in a Huankai hand carrying case, with a
sampling head, calibration certificate, manual, warranty and
battery charger.
Microbial Air Sampler