1.adopts the fluorescence spectroscopy principle 2.quick, effective and easy data collection

Applications

Model TH-2002 UV fluorescence method SO2 analyzer, which is one of the high technology

products of ambient air automatic monitoring, adopts the fluorescence spectroscopy principle,

applies the modern technology of micro-SCM to detect and control signals, processes signal

data and etc. It uses optics, mechanism, electron, photochemistry and modern sensor technology 

synthetically which can continuous monitor the concentration of  the SO2 in ambient.

 Operation Principle: Zinc lamp can emit ultraviolet light, which is filtered and becomes monochromatic light (wavelength is 214nm) which then focus on the SO2 reaction chamber. Sample SO2 which absorbs ultraviolet light can make energy level transition, so SO2 becomes ground state into excitation state. The SO2 in excitation state is not stable and emits fluorescence with 330nm centre wavelength. The degree of the fluorescence light is directly related to SO2 concentration. The optical signal is changed into electric signal by PMT, through which SO2 concentration is measured. To avoid error made by fluctuation of light path, where there is two paths monitoring, the one is fluorescence intensity which is excited by light path measuring of sample gas, the other is reference light path to monitor the change of Zinc lamp light source and make auto-compensation to the change of light source.

  Features

auto alarm indicator facilitates fault determination and decreases failure rate.Modular design is easy for daily maintenance and repair.Quick, effective and easy data collection

Specifications

NO.

Item Name

Technical indicator

1

Measuring range

(0~0.5)ppm or(0~1.0)ppm(can be extended)

2

Resolution

0.1ppb

3

Noise

Zero point noise

1ppb

Span noise 

5ppb

4

Minimum detectable value

2ppb

5

Stability

Zero drift

±5ppb/24hr

Span

drift

20% Span drift

±3ppb/24hr

80% Span drift

±5ppb/24hr

6

Precision

20% measuring span 

±3ppb

80% measuring span

±5ppb

7

Linearity

Related coefficient(r)

more than 0.999

Slope rate(a)

0.99-1.01

intercept(b)

less than 1%F.S

8

Voltage stability

±1%F.S

9

Response time

Rising response time(T90)

3min(Only the analyzer itself, the gas 

distributing system response time not

included)

Falling response time(T10)

3min(see the above description)